USER’S MANUAL
GENES
I
S 2002
TM
GENES
I
S 2006
TM
PLASMA THAWING WATER BATH SYSTEM
GenesisBPS, LLC
65 Commerce Way
Hackensack, NJ 07601 U.S.A.
Tel: 201-708-1400
866-712-5663
Fax: 201-704-1104
www.genesisbps.com
0609
Vista de página 0
1 2 3 4 5 6 ... 21 22

Resumo do Conteúdo

Página 1

USER’S MANUAL GENESIS 2002 TM GENESIS 2006 TM PLASMA THAWING WATER BATH SYSTEM GenesisBPS, LLC 65 Commerce W

Página 2 - GENESIS 2006 SHOWN

5. TO CLEAN THE BATH OR TO ADD WATER SAFETY : NEVER ADD WATER OR MOVE THE BATH IF THE POWER SWITCH (ON THE LEFT SIDE PANEL)

Página 3 - 866-712-5663

when filled with water. Do not drop the bath on a hard surface. GENESIS 2006: Remove all, or part of the water. The bath weighs approximately

Página 4

6. CONTROL PANEL AND INDICATORS GENESIS 2002TM CONTROL PANEL 12

Página 5

GENESIS 2006TM CONTROL PANEL 13

Página 6

CONTROL KEYS POWER ... This key turns the unit on and off when the POWER SWITCH on the left panel is turned ON. The POWER S

Página 7

INDICATORS HEATING ... This red indicator is lit when the heating element is actually drawing current from the circuit and hea

Página 8

2) The ADD WATER light is on and there is sufficient water in the bath. • With the unit unplugged and raised out of the water, check that

Página 9

7) Pump does not circulate the water. • Debris may be in the pump impeller housing. Raise the shroud unit and lock in the up position.

Página 10

8. SPECIAL FEATURES • Thaws multiple plasma units simultaneously • Thaws standard size plasma as well as cryoprecipitate • Individual time settin

Página 11

10. SPECIFICATIONS GENESIS 2002TM GENESIS 2006TM External Dimension 21.6” L x 8.3” Wide x 13.2” High 26.6” L x 9.6” Wide x 13.2” H

Página 12

GENESIS 2006 SHOWN 2

Página 13

11. ORDERING/RETURN INFORMATION ORDERING INFORMATION : To place an order for a part or an accessory, please call : GenesisBPS, LLC 6

Página 14

Hospital/Account Name Department and shipping address Name and telephone number of contact Date GENESIS 2002/2006

Página 15

GENESIS 2002 GENESIS 2006 PLASMA THAWING WATER BATH SYSTEM WARRANTY The GENESIS 2002/2006, Plasma Thawing Water Bath System is warranted a

Página 16 - GenesisBPS

To assure proper record keeping, please retain this information for future use. DATE PURCHASED ____________________________ SERIAL # __________

Página 17

CONTENTS page 1. UNPACKING ... 5 2. TO THAW BLOOD PROD

Página 18

1. UNPACKING 1.1 The following is supplied with your GENESIS 2002/2006 unit. Control Unit Water Bath Thawing B

Página 19

indicator will light. To correct this situation, fill bath with water as explained above. Add 2 ml of LABath CLEANTM each time the water is ch

Página 20

c. To change the default SET TEMP : You can adjust this temperature within the range of ambient to + 56 °C, in one degree increments. To adjust, p

Página 21 - GenesisBPS, LLC. (as above)

thawing time may vary. The amount of air left in the plasma thawing bag during thawing will also affect thawing time. The laboratory personnel ca

Página 22 - WARRANTY

3. TO REMOVE A THAWED BAG 3.1 Press the ROCK key to stop the rocking motion. a. Remove the thawed bag (and any weights used) from the P